Res Med Recalls CPAP Machines

Florence Cardinal Health Guide
  • Res Med, one of the world's largest suppliers of CPAP (Continuous Positive Airway Pressure} machines, used in the treatment of sleep apnea, today announced a worldwide voluntary recall of approximately 300,000 of its early production S8 flow generators. The recall was prompted by the remote possibility of a short circuit in the power supply connector.

     

    Res Med states: Patients may continue to use their S8 flow generators until they receive a replacement device. As with any electrical device, patients should make sure that it is placed on a hard clean surface and that the area around the device is clear during use. Patients should discontinue use of the device if there are any signs of electrical failure such as intermittent power, cracking sounds, sparking or charred smell. Patients should not use supplemental oxygen with an affected device; patients using supplemental oxygen should immediately contact their home healthcare provider for a replacement.

    Add This Infographic to Your Website or Blog With This Code:

     

    In only seven cases, less than 2/10th of 1% of the machines in operation, has this power component caused a problem. However, because of the potential fire hazard, all machines are being recalled. Serial numbers of machines to be recalled are printed on the base of the machines. Serial numbers include the following:

     

    From

    To

    20040285613

    20060269563

    20060275728

    20060276751

    20060277160

    20060277415

    20060281672

    20060281991

    20060283424

    20060283743

    20060284896

    20060285445

    20060287568

    20060290823

    20060292360

    20060294694

    20060312361

    20060312597

    20060318692

    20060319459

    20060325074

    20060327794

    20060330588

    20060331043

Published On: May 02, 2007